In the last decades a great number of in vitro studies have been conducted to improve the design and to understand the transvalvular flow patterns under steady-state and pulsating flow conditions. Steady-state tests are useful for studying the flow established upstream and downstream of the valve prosthesis in different flow conditions and, in particular, at the peak flow rate. In the present study, the particle image velocimetry (PIV) technique is employed to visualize the flow patterns in a precommercial model of a bi-leaflet mechanical heart valve prosthesis in a steady-state flow regime. The use of the PIV technique and a convenient test rig provide good conditions with which to investigate the whole flow field upstream and downstream of the valve. Velocity and vorticity maps are obtained for the flow passing through the prosthesis in different axial planes. A new and simple solution for the measurement test chamber is proposed. This innovative approach of observing the flow in staggered planes (other than diametrical planes) allows a flow analysis to be performed upstream and downstream of the valve in regions near the tube wall where it has the maximum potential for injury of the vessel's inner cell layer.

Particle Image Velocimetry for flow analysis in longitudinal planes across a mechanical artificial heart valve / Castellini, Paolo; Pinotti, M.; Scalise, Lorenzo. - In: ARTIFICIAL ORGANS. - ISSN 0160-564X. - STAMPA. - 28:5(2004), pp. 507-513. [10.1111/j.1525-1594.2004.07271.x]

Particle Image Velocimetry for flow analysis in longitudinal planes across a mechanical artificial heart valve

CASTELLINI, Paolo;SCALISE, Lorenzo
2004-01-01

Abstract

In the last decades a great number of in vitro studies have been conducted to improve the design and to understand the transvalvular flow patterns under steady-state and pulsating flow conditions. Steady-state tests are useful for studying the flow established upstream and downstream of the valve prosthesis in different flow conditions and, in particular, at the peak flow rate. In the present study, the particle image velocimetry (PIV) technique is employed to visualize the flow patterns in a precommercial model of a bi-leaflet mechanical heart valve prosthesis in a steady-state flow regime. The use of the PIV technique and a convenient test rig provide good conditions with which to investigate the whole flow field upstream and downstream of the valve. Velocity and vorticity maps are obtained for the flow passing through the prosthesis in different axial planes. A new and simple solution for the measurement test chamber is proposed. This innovative approach of observing the flow in staggered planes (other than diametrical planes) allows a flow analysis to be performed upstream and downstream of the valve in regions near the tube wall where it has the maximum potential for injury of the vessel's inner cell layer.
2004
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11566/32558
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 14
  • ???jsp.display-item.citation.isi??? 13
social impact