Introduction Controlling temperature generation during implant site preparation is important to prevent implant early failure. Aim The present in vitro study aimed at measuring temperature variation generated during the initial osteotomy using both rotatory and piezo-surgical inserts. Methods Nine groups were defined according to drill and insert type, cooling volume (mL/min) and cooling temperature. A total of 315 implant site preparations were performed in an artificial bone sample and the temperature was measured using an infrared camera. Drills' wear was assessed using scanning electron microscopy at baseline and after 10 and 35 utilizations. Results Piezo-surgical insert groups determined a temperature increase that was significantly higher than the one generated by rotatory drills groups (p < 0.001). When considering rotatory drills groups a temperature >= 40 degrees C was never recorded. Conclusion Lower saline temperature implied a significant temperature decrease (p < 0.001), while the increase in cooling volume did not imply a temperature decrease. The scanning electron microscopy analysis of the drills demonstrated that little drill wear occurred up to 35 utilization times.

Heat Generation During Initial Osteotomy for Implant Site Preparation: An In Vitro Measurement Study / Aquilanti, Luca; Antognoli, Luca; Rappelli, Giorgio; Di Felice, Roberto; Scalise, Lorenzo. - In: JOURNAL OF MAXILLOFACIAL & ORAL SURGERY. - ISSN 0972-8279. - ELETTRONICO. - 22:2(2023), pp. 313-320. [10.1007/s12663-022-01800-8]

Heat Generation During Initial Osteotomy for Implant Site Preparation: An In Vitro Measurement Study

Aquilanti, Luca;Antognoli, Luca;Rappelli, Giorgio;Scalise, Lorenzo
2023-01-01

Abstract

Introduction Controlling temperature generation during implant site preparation is important to prevent implant early failure. Aim The present in vitro study aimed at measuring temperature variation generated during the initial osteotomy using both rotatory and piezo-surgical inserts. Methods Nine groups were defined according to drill and insert type, cooling volume (mL/min) and cooling temperature. A total of 315 implant site preparations were performed in an artificial bone sample and the temperature was measured using an infrared camera. Drills' wear was assessed using scanning electron microscopy at baseline and after 10 and 35 utilizations. Results Piezo-surgical insert groups determined a temperature increase that was significantly higher than the one generated by rotatory drills groups (p < 0.001). When considering rotatory drills groups a temperature >= 40 degrees C was never recorded. Conclusion Lower saline temperature implied a significant temperature decrease (p < 0.001), while the increase in cooling volume did not imply a temperature decrease. The scanning electron microscopy analysis of the drills demonstrated that little drill wear occurred up to 35 utilization times.
2023
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11566/317332
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 1
social impact