Purpose: Contact with cardiac tissue is a determinant of lesion efficacy during atrial fibrillation (AF) ablation. The Sensei®X Robotic Catheter System (Hansen Medical, CA) has been validated for contact force sensing. The electrical coupling index (ECI) from the EnSite Contact™ system (St. Jude Medical, MN) has been validated as an indicator of tissue contact. We aimed at analyzing ECI behavior during radiofrequency (RF) pulses maintaining a stable contact through the robotic navigation contact system. Methods: In 15 patients (age, 59 ± 12) undergoing AF ablation, pulmonary vein (PV) isolation was guided by the Sensei®X System, employing the Contact™ catheter. Results: During the procedure, we assessed ECI changes associated with adequate contact based on the IntelliSense® force-sensing technology (Hansen Medical, CA. Baseline contact (27 ± 8 g/cm2) ECI value was 99 ± 13, whereas ECI values in a noncontact site (0 g/cm2) and in a light contact site (1-10 g/cm2) were respectively 66 ± 12 and 77 ± 10 (p < 0.0001). Baseline contact ECI values were not different depending on AF presentation (paroxysmal AF, 98 ± 9; persistent AF, 100 ± 9) or on cardiac rhythm (sinus rhythm, 97 ± 7; AF,101 ± 10). In all PVs, ECI was significantly reduced during and after ablation (ECI during RF, 56 ± 15; ECI after RF, 72 ± 16; p < 0.001). A mean reduction of 32.2 % during RF delivery and 25.4 % immediately after RF discontinuation compared with baseline ECI was observed. Conclusions: Successful PV isolation is associated with a significant decrease in ECI of at least 20 %. This may be used as a surrogate marker of effective lesion in AF ablation. © 2014 Springer Science+Business Media.
Simultaneous assessment of contact pressure and local electrical coupling index using robotic navigation / Dello Russo, A.; Fassini, G.; Casella, M.; Bologna, F.; Al-Nono, O.; Colombo, D.; Biagioli, V.; Santangeli, P.; Di Biase, L.; Zucchetti, M.; Majocchi, B.; Marino, V.; Gallinghouse, J. J.; Natale, A.; Tondo, C.. - In: JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY. - ISSN 1383-875X. - 40:1(2014), pp. 23-31. [10.1007/s10840-014-9882-2]
Simultaneous assessment of contact pressure and local electrical coupling index using robotic navigation
Dello Russo A.;Casella M.;
2014-01-01
Abstract
Purpose: Contact with cardiac tissue is a determinant of lesion efficacy during atrial fibrillation (AF) ablation. The Sensei®X Robotic Catheter System (Hansen Medical, CA) has been validated for contact force sensing. The electrical coupling index (ECI) from the EnSite Contact™ system (St. Jude Medical, MN) has been validated as an indicator of tissue contact. We aimed at analyzing ECI behavior during radiofrequency (RF) pulses maintaining a stable contact through the robotic navigation contact system. Methods: In 15 patients (age, 59 ± 12) undergoing AF ablation, pulmonary vein (PV) isolation was guided by the Sensei®X System, employing the Contact™ catheter. Results: During the procedure, we assessed ECI changes associated with adequate contact based on the IntelliSense® force-sensing technology (Hansen Medical, CA. Baseline contact (27 ± 8 g/cm2) ECI value was 99 ± 13, whereas ECI values in a noncontact site (0 g/cm2) and in a light contact site (1-10 g/cm2) were respectively 66 ± 12 and 77 ± 10 (p < 0.0001). Baseline contact ECI values were not different depending on AF presentation (paroxysmal AF, 98 ± 9; persistent AF, 100 ± 9) or on cardiac rhythm (sinus rhythm, 97 ± 7; AF,101 ± 10). In all PVs, ECI was significantly reduced during and after ablation (ECI during RF, 56 ± 15; ECI after RF, 72 ± 16; p < 0.001). A mean reduction of 32.2 % during RF delivery and 25.4 % immediately after RF discontinuation compared with baseline ECI was observed. Conclusions: Successful PV isolation is associated with a significant decrease in ECI of at least 20 %. This may be used as a surrogate marker of effective lesion in AF ablation. © 2014 Springer Science+Business Media.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.