In this paper we present a novel, non-invasive measurement system for optical monitoring of the cardiac rate. This measurement method, called vibrocardiography (VCG), is based on the use of a laser Doppler vibrometer for the measurement of the velocity of displacement of the skin in correspondence of the chest wall. We report the typical VCG signals measured on 4 healthy subject and in particular, we investigated the effect on the signal of the measurement position respect to the chest wall as well as the effect of the surface characteristics on the measured signal.
A laser Doppler approach to cardiac motion monitoring: Effects of surface and measurement position / Scalise, Lorenzo; U., Morbiducci; M., De Melis. - 6345/2006:(2006), pp. 63450D.1-63450D.1-63450D.1-63450D.11. (Intervento presentato al convegno 7th International Conference on Vibration Measurements by Laser Techniques tenutosi a Ancona nel 19-22 June 2006) [10.1117/12.69315].
A laser Doppler approach to cardiac motion monitoring: Effects of surface and measurement position
SCALISE, Lorenzo;
2006-01-01
Abstract
In this paper we present a novel, non-invasive measurement system for optical monitoring of the cardiac rate. This measurement method, called vibrocardiography (VCG), is based on the use of a laser Doppler vibrometer for the measurement of the velocity of displacement of the skin in correspondence of the chest wall. We report the typical VCG signals measured on 4 healthy subject and in particular, we investigated the effect on the signal of the measurement position respect to the chest wall as well as the effect of the surface characteristics on the measured signal.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.