This paper reports a novel technique to estimate the total averaged power emitted by a device under test placed inside the working volume of a reverberation chamber in a radiated emission test. The electromagnetic field inside the chamber is stirred by the application of the multiple monopole source stirring in receiving mode, that is a particular way to implement the source stirring technique; in this scenario an array of monopole antenna is placed onto the chamber walls and sequentially connected to the measurement system to get the stirring action. The procedure for the radiated emission estimation requires that chamber has to be calibrated by a set of measurements. In the paper the calibration results are compared to the prediction of the analytical model, developed to predict the chamber's performances.

Calibration of a Reverberation Chamber for Radiated Emission Measurements by Wall Mounted Antennas / De Leo, A.; Cerri, G.; Russo, P.; Primiani, V. M.. - ELETTRONICO. - 2019-:(2019), pp. 1893-1899. (Intervento presentato al convegno 2019 PhotonIcs and Electromagnetics Research Symposium - Spring, PIERS-Spring 2019 tenutosi a ita nel 2019) [10.1109/PIERS-Spring46901.2019.9017542].

Calibration of a Reverberation Chamber for Radiated Emission Measurements by Wall Mounted Antennas

De Leo A.
;
Cerri G.;Russo P.;Primiani V. M.
2019-01-01

Abstract

This paper reports a novel technique to estimate the total averaged power emitted by a device under test placed inside the working volume of a reverberation chamber in a radiated emission test. The electromagnetic field inside the chamber is stirred by the application of the multiple monopole source stirring in receiving mode, that is a particular way to implement the source stirring technique; in this scenario an array of monopole antenna is placed onto the chamber walls and sequentially connected to the measurement system to get the stirring action. The procedure for the radiated emission estimation requires that chamber has to be calibrated by a set of measurements. In the paper the calibration results are compared to the prediction of the analytical model, developed to predict the chamber's performances.
2019
978-1-7281-3403-1
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11566/276239
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact